Research and Development of a Breathable Multimodal Skin Patch for Sleep Monitoring: Anatomical, Materials, and Computational Foundations for Concurrent EEG, ECG, and EDA Recording

Abstract

Laboratory polysomnography (PSG), the reference standard for measuring sleep, requires many wired sensors and an unfamiliar setting, which makes it uncomfortable for patients and impractical across many nights. Consumer wearables are comfortable but do not record brain activity and classify individual sleep stages, particularly rapid eye movement (REM) sleep, with limited accuracy. A thin, breathable skin patch that records electroencephalography (EEG), electrocardiography (ECG), and electrodermal activity (EDA) from one discreet site could bridge this gap. Such a device does not yet exist, and several foundational questions must be answered before one can be built. This proposal describes the initial research and development (R&D) phase, organized around three fronts. Aim 1 (neurophysiology and anatomy) will determine how the anatomy and skin properties of the upper neck and retroauricular region constrain signal quality, using volume-conductor modeling and small-scale measurements with conventional electrodes to identify the electrode layout that maximizes the quality of all three signals. Aim 2 (materials science) will identify breathable substrates, electrodes, and adhesives that are safe for prolonged skin contact and electrically stable, culminating in small-format test patches recorded alongside conventional electrodes. Aim 3 (machine learning) will build the quantitative pipeline that scores signal quality, separates artifacts, and tests how much sleep-relevant information, including REM-related features, can be extracted from the signals available at this site. The expected outcome is an evidence-based design specification for a first prototype, together with a clear account of anatomical and technical limits, to support later development for REM sleep tracking and for more comfortable clinical sleep studies.

Specific Aims

Sleep is organized into recurring cycles of non-rapid eye movement (NREM) and REM sleep, each defined by distinct cortical rhythms and accompanied by systematic shifts in autonomic tone. Measuring sleep accurately therefore requires recording both the brain and the body. PSG does this comprehensively, but its cost, many wired sensors, and laboratory setting make it burdensome for patients and limit it to one or two nights. Consumer wearables are comfortable but infer sleep from movement and optical heart rate and agree less closely with PSG when classifying individual sleep stages (de Zambotti et al., 2019; Chinoy et al., 2021). Prior work has shown that sleep-relevant EEG can be recorded from behind the ear (Debener et al., 2015; Mikkelsen et al., 2017; Sterr et al., 2018) and that thin, gas-permeable electronics can be worn on the skin with minimal irritation (Kim et al., 2011; Miyamoto et al., 2017). No device yet combines cortical, cardiac, and sympathetic recording at one comfortable site.

Before such a patch can be built, three foundational questions must be answered: whether the anatomy of a single site allows all three signals to be recorded, which materials can be worn safely and stably through the night, and how the resulting data can be quantified and interpreted. The long-term goal of this research is a comfortable, minimally obtrusive patch for multinight sleep measurement, including REM sleep tracking, in research and clinical settings. The objective of this R&D phase is to establish the anatomical, materials, and computational foundations needed to design a prototype. The central hypothesis is that a continuous region spanning the upper neck and the area behind the ear can support recording of EEG and ECG of usable quality, with EDA feasibility determined empirically, using breathable, skin-safe materials. This hypothesis will be tested through three specific aims.

Aim 1. Neurophysiology and anatomy

Determine how the anatomy of the retroauricular–cervical region constrains signal quality, and identify the electrode layout that maximizes it. Volume-conductor modeling and a small-scale measurement study with conventional electrodes will compare candidate layouts against reference EEG, ECG, and EDA. Electrode spacing and orientation, reference and ground placement, muscle interference, hair, and skin properties will be varied or measured.

Working hypothesis: at least one layout will meet pre-specified quality criteria for EEG and ECG simultaneously.

Aim 2. Materials science

Identify materials that allow a breathable, skin-safe, electrically stable patch. Candidate porous substrates, electrode materials, and skin adhesives will be screened on the bench for breathability, flexibility, chemical stability, and biocompatibility, then tested on skin. Shortlisted materials will be combined into small-format test patches recorded alongside conventional electrodes.

Working hypothesis: a porous substrate with a suitable electrode–adhesive combination will maintain stable electrode–skin contact over 8 h without skin irritation.

Aim 3. Machine learning

Build a pipeline that quantifies the measurements and estimates the sleep information available at this site. Signal-quality scoring, artifact separation, and feature extraction will be developed for the data from Aims 1 and 2. Using public PSG datasets, models restricted to the signals a patch could record will estimate how well sleep stages, including REM sleep, can be identified.

Working hypothesis: combining EEG with cardiac and muscle-tone features will improve REM sleep identification over EEG alone.

Expected outcome

The expected outcome is an evidence-based design specification for a first prototype, including electrode layout, materials, and analysis pipeline, together with a documented account of the anatomical and technical limits of single-site recording. These results will set up subsequent prototype development and overnight validation against PSG, aimed at REM sleep tracking and more comfortable clinical sleep studies.

Background and Significance

Sleep as a coordinated brain–body state

Under the American Academy of Sleep Medicine (AASM) rules, sleep is scored in 30-s epochs as wake (W), NREM stages N1, N2, and N3, or REM (R), using EEG, electrooculography (EOG), and chin electromyography (EMG) (Berry et al., 2020). Each stage has characteristic cortical signatures: posterior alpha activity (8–12 Hz) during relaxed wakefulness, sleep spindles (11–16 Hz) and K-complexes in N2, high-amplitude slow waves (0.5–2 Hz) in N3, and low-amplitude mixed-frequency activity accompanied by rapid eye movements and muscle atonia in REM sleep. The autonomic nervous system changes in parallel. Sympathetic outflow declines and parasympathetic (vagal) influence on heart rate increases with deepening NREM sleep, whereas REM sleep is marked by sympathetic surges and greater cardiovascular variability (Somers et al., 1993; Trinder et al., 2001). Electrodermal activity follows a distinct pattern: bursts of skin conductance responses occur most frequently during slow-wave sleep and are comparatively rare during REM sleep (Boucsein, 2012; Sano et al., 2014). Because cortical and autonomic signatures carry partially independent information, combining them may resolve stages that are ambiguous from any single signal. REM sleep is of particular interest to theories linking sleep physiology to dreaming and memory (Hobson et al., 2000), yet its identification depends on features, such as eye movements and muscle atonia, that are difficult to measure outside the laboratory.

Limitations of current sleep measurement

PSG remains the reference standard, but it requires trained technicians, many wired sensors attached to the scalp, face, chest, and limbs, and an unfamiliar environment that can itself alter sleep. For patients, particularly children, older adults, and people with anxiety or sensory sensitivities, this setup can be uncomfortable and can degrade the very sleep being measured. It is also rarely repeated across nights. Consumer devices offer the opposite trade-off. Validation studies indicate that wrist-worn and other consumer sleep trackers generally classify sleep versus wake well but are less accurate in estimating wake after sleep onset and in discriminating individual sleep stages (de Zambotti et al., 2019; Chinoy et al., 2021). Because these devices do not record cortical activity, their staging relies on indirect correlates such as movement and heart rate.

Anatomical considerations at the retroauricular–cervical site

Several groups have shown that EEG can be recorded from in and around the ear (Looney et al., 2012; Debener et al., 2015), and behind-the-ear electrode arrays have supported automatic sleep staging with moderate-to-substantial agreement with PSG-based scoring (Mikkelsen et al., 2017; Sterr et al., 2018). The region spanning the upper neck and the area behind the ear is anatomically complex. It overlies the mastoid process and the temporal and occipital bones, which attenuate and spatially blur cortical signals; it lies close to the sternocleidomastoid, trapezius, and posterior auricular muscles, whose activity produces EMG with a broad spectrum that overlaps substantially with EEG and is most prominent at temporal sites (Goncharova et al., 2003); it varies in hair coverage, skin thickness, and sebaceous activity across individuals; and it is subject to pressure and shear when a person lies on a pillow. Each of these factors affects signal quality and electrode contact. Some may also be informative: loss of muscle tone is a defining feature of REM sleep, so sensitivity to cervical EMG may carry information useful for REM detection if cortical and muscular sources can be separated.

Cardiac and electrodermal measures, and the skin interface

Heart rate variability (HRV) provides a noninvasive index of autonomic regulation. Time-domain indices such as the root mean square of successive differences (RMSSD) and frequency-domain indices such as high-frequency (HF; 0.15–0.40 Hz) power primarily reflect vagal modulation, whereas low-frequency (LF; 0.04–0.15 Hz) power reflects a mixture of sympathetic and parasympathetic influences (Task Force of the ESC and NASPE, 1996; Shaffer and Ginsberg, 2017). Accurate HRV analysis requires precise beat-to-beat (R–R) intervals, best obtained from ECG. EDA reflects eccrine sweat gland activity under sympathetic cholinergic control and therefore complements HRV as a sympathetic measure (Boucsein, 2012). Most EDA research uses the palms or fingers, where eccrine gland density is high; the neck and retroauricular regions have lower gland density, so EDA feasibility at this site is treated as an open question. All three measurements also depend on the biochemistry of the electrode–skin interface. The stratum corneum, its hydration, and the electrolyte content of sweat determine electrode–skin impedance and electrode potentials, and EDA in particular requires nonpolarizing electrodes and a stable electrolyte interface (Boucsein, 2012).

Materials for long-term skin contact

Epidermal electronics established that ultrathin, stretchable devices can conform to the skin and record electrophysiological signals with mechanical properties similar to those of skin (Kim et al., 2011). Subsequent work demonstrated gas-permeable nanomesh conductors that can be worn for extended periods with minimal inflammatory response, highlighting the importance of breathability for long-term wear (Miyamoto et al., 2017). Occlusive materials trap sweat and moisture, which can degrade adhesion, alter impedance, and irritate skin over a full night. For a sleep patch, breathability and biocompatibility are therefore requirements for both safety and signal stability. More broadly, recent clinical work on AI-integrated wearable sensors, such as wearable ultrasound for noninvasive estimation of central venous pressure, illustrates the path from sensor design to clinical validation that this project aims to follow (Dong et al., 2026).

Machine learning for wearable sleep data

Deep learning models trained on one or a few EEG channels can approximate expert sleep scoring (Supratak et al., 2017; Perslev et al., 2021). Two issues are critical for wearable data. First, wearable recordings contain artifacts from movement, muscle activity, and changing electrode contact that are uncommon in curated PSG datasets. Second, models frequently perform worse on individuals outside the training set, so evaluation must be strictly subject-independent. Large public PSG datasets (Kemp et al., 2000; Goldberger et al., 2000; Zhang et al., 2018) make it possible to estimate, before a device exists, how much sleep information is retained when only a restricted set of signals is available.

Significance

This R&D phase addresses the questions that must be answered before a multimodal sleep patch can be designed responsibly. The results have value regardless of outcome: a positive result provides a design specification grounded in anatomy, materials testing, and quantitative analysis, whereas a negative or partial result defines the physical limits of single-site recording and indicates which signals require separate placement. In the longer term, a validated patch could support REM sleep tracking across many nights and make clinical sleep studies more comfortable and accessible by reducing the number of wired sensors patients must wear.

Innovation

  • Anatomy-driven design. Electrode placement is treated as an experimental question grounded in anatomy, skin properties, and volume-conductor modeling, rather than assumed.

  • Single-site multimodal recording. The project tests whether cortical, cardiac, and sympathetic signals can be recorded together from one discreet site.

  • Muscle activity treated as both artifact and signal. Cervical EMG is separated from EEG and evaluated as a potential index of muscle tone for REM detection.

  • Breathability and biocompatibility as design requirements. Materials are selected for skin safety and full-night electrical stability from the outset.

  • Computational feasibility before fabrication. Public PSG data are used to estimate the sleep information available from patch-compatible signals before a full device is built.

Research Design and Methods

Overview

The three aims run in parallel and inform one another (Table 2). Aim 1 defines where and how electrodes should be placed. Aim 2 identifies the materials from which a patch can be made and produces small-format test patches. Aim 3 provides the analysis pipeline for the data from Aims 1 and 2 and estimates, using public datasets, how much sleep information the selected signals can carry. The first manuscript will combine a literature review of candidate anatomical sites with the Aim 1 modeling results. This phase does not build or validate a complete overnight device; that work is described under Future Directions.

Participants

Healthy adults aged 18–45 years will be recruited. Exclusion criteria are a diagnosed sleep disorder, cardiac arrhythmia, or neurological disorder; use of medications known to affect sleep or autonomic function; known skin disease at the test sites; and known sensitivity to medical adhesives. Habitual sleep quality will be assessed with the Pittsburgh Sleep Quality Index (Buysse et al., 1989). Recruitment will aim for balance across sexes and will include a range of skin types and hair characteristics, because these affect electrode contact at the retroauricular site. Planned sample sizes are 20 participants for the Aim 1 measurement study, 10 for Aim 2 skin-wear testing, and 8 for the Aim 2 test-patch pilot. Because this is an R&D phase, sample sizes are chosen to inform estimation and design decisions rather than to confirm hypotheses.

Aim 1: Neurophysiology and Anatomy

The candidate site must serve three signals with competing requirements. EEG benefits from proximity to cortical generators and isolation from muscle; ECG benefits from wide electrode spacing aligned with the cardiac electrical axis; and EDA requires adequate sweat gland density and a stable, nonpolarizing electrode interface. Aim 1 determines whether, and in which configuration, the region spanning the upper neck and the area behind the ear satisfies these requirements.

Aim 1a: Anatomical review and modeling

A structured literature review will summarize the anatomy of the region (bone, muscle, hair coverage, skin thickness, and sweat and sebaceous gland distribution) and published recordings of ear-EEG, neck-based ECG, and nonpalmar EDA, extracting signal amplitudes, signal-to-noise ratios, and artifact profiles by site. A realistic standardized head model (Huang et al., 2016) will be used to compute lead fields for candidate bipolar electrode pairs within the region and to estimate their sensitivity to cortical sources relevant to sleep, including frontocentral generators of slow waves and spindles and posterior generators of alpha activity. The output is a ranked shortlist of three to five candidate layouts.

Aim 1b: Small-scale measurement study

  • Design. Each participant will wear a dense grid of conventional Ag/AgCl electrodes over the candidate region on one side, allowing the shortlisted layouts to be reconstructed by re-referencing. Reference signals will be recorded concurrently: scalp EEG at standard 10–20 positions (Klem et al., 1999) referenced to the contralateral mastoid, chest-lead ECG, and palmar EDA measured exosomatically with a constant-voltage method. At each grid location, skin properties relevant to the electrode interface will be documented, including hair coverage, skin hydration, and baseline electrode–skin impedance.

  • Protocol. Each session will include (a) resting recordings with eyes open and eyes closed to assess alpha reactivity; (b) controlled artifact tasks, including jaw clenching, swallowing, head rotation, and lying on each side with pillow pressure on the site; (c) sympathetic challenges known to elicit skin conductance responses, including deep inspiration, brief auditory tones, and mental arithmetic; (d) paced breathing to assess respiratory modulation of heart rate; and (e) a 90-min afternoon nap opportunity to capture sleep onset, spindles, and, where present, slow waves.

  • Analysis. Signal processing for this aim is carried out with the pipeline developed in Aim 3. R-peaks will be detected with an established QRS detection algorithm (Pan and Tompkins, 1985), and phasic and tonic EDA components will be separated using continuous decomposition analysis (Benedek and Kaernbach, 2010). Agreement with reference measures will be summarized with Bland–Altman analysis (Bland and Altman, 1986) and intraclass correlation coefficients. Effects of layout and skin properties will be estimated with linear mixed-effects models with participant as a random intercept, with false discovery rate correction across outcomes.

  • Quality criteria. Provisional criteria are listed in Table 1 and will be finalized after Aim 1a and preregistered before data collection. A layout is carried forward if it meets the EEG and ECG criteria; EDA performance is a secondary criterion.

Table 1. Provisional signal-quality criteria for candidate electrode layouts (Aim 1b).
SignalMeasureProvisional criterion
EEGAlpha reactivity (eyes closed vs. eyes open, 8–12 Hz power)Significant within-participant increase with eyes closed
EEGSpindle detection agreement with reference central EEGF1 score ≥ 0.60
EEGEpoch-wise spectral correlation with reference EEG, 0.5–16 HzMedian r ≥ 0.70
ECGR-peak detection vs. chest-lead ECGSensitivity and positive predictive value ≥ 99%
ECGR–R interval error; RMSSD agreementMean absolute error ≤ 5 ms; ICC ≥ 0.90
EDADetection of palmar skin conductance responses ≥ 0.05 µSSensitivity ≥ 50% (secondary criterion)
ArtifactEMG contamination and recovery during artifact tasksReported descriptively; used to rank layouts

Expected outcomes and alternative approaches

Based on prior behind-the-ear recordings, EEG of usable quality is expected for at least one layout. ECG amplitude at the neck is expected to be lower than at the chest; if R-peak criteria are not met, electrode spacing will be extended inferiorly along the neck to better align with the cardiac axis. EDA is the signal at greatest risk because of lower sweat gland density. If EDA criteria are not met, two alternatives will be evaluated: a small satellite EDA electrode on a higher-density site connected to the main patch, or a design that omits EDA. If muscle contamination limits EEG quality, analyses will emphasize frequencies below 16 Hz and use EMG-referenced artifact regression.

Aim 2: Materials Science

A patch worn through the night must remain breathable, flexible, adherent, and electrically stable, and it must be safe for repeated skin contact. Aim 2 identifies materials that meet these requirements and produces small-format test patches, without yet building a complete device.

Aim 2a: Bench screening

  • Substrates. Candidate porous substrates, including electrospun polymer nanofiber membranes, porous silicone elastomers, and microperforated medical-grade polyurethane films, will be characterized for thickness, stretchability, elastic modulus, and water vapor transmission rate measured according to ASTM E96.

  • Electrodes. Thin-film Ag/AgCl, gold, and conductive-polymer electrodes will be compared, with and without a thin hydrogel interface. Because exosomatic EDA requires nonpolarizing electrodes (Boucsein, 2012), Ag/AgCl will be prioritized for EDA contacts. Electrodes will be tested for impedance spectra (1 Hz–1 kHz), electrode noise, and DC potential drift, and for stability after soaking in artificial sweat.

  • Adhesives. Skin-contact adhesives will be evaluated for peel strength under dry and sweat-exposed conditions and for chemical compatibility with the electrode materials, including corrosion or delamination after prolonged contact.

  • Biocompatibility. Only materials with existing skin-contact biocompatibility data will be advanced to human testing, and candidate assemblies will be evaluated against the relevant parts of the ISO 10993 series for skin-contacting devices (cytotoxicity, irritation, and sensitization) before wear testing.

Aim 2b: Skin-wear testing

Ten participants will wear small samples of shortlisted materials, without electronics, on the retroauricular–cervical site for up to 8 h. Outcomes will include standardized visual scoring of erythema and edema before and after wear, transepidermal water loss, a comfort rating on a visual analog scale, and the proportion of sample area remaining adherent at removal.

Aim 2c: Small-format test patches

Shortlisted substrates, electrodes, and adhesives will be combined into small-format test patches implementing the layout selected in Aim 1. Eight participants will wear the test patches alongside conventional electrodes during resting tasks and a nap, using a tethered laboratory amplifier. Signals will be compared with the conventional electrodes using the Aim 1 criteria, and electrode–skin impedance will be tracked across the session.

Success criteria

A material combination is carried forward if impedance at 10 Hz increases by less than 50% from baseline over 8 h of wear, at least 90% of the sample remains adherent at removal, no reaction exceeds minimal erythema, comfort is not inferior to standard PSG electrodes, and test-patch signals meet the Aim 1 EEG and ECG criteria. If dry electrodes show excessive impedance, a thin hydrogel interface will be adopted; if adhesion is insufficient, a breathable overlay will be evaluated.

Aim 3: Machine Learning

Measurements from Aims 1 and 2 must be quantified consistently, and the sleep information available at the patch site must be estimated before committing to a full device. Aim 3 provides both.

Aim 3a: Signal-quality and artifact pipeline

A shared processing pipeline will be developed for all recordings in this phase. It will include band-pass and notch filtering; automated signal-quality indices for each signal and epoch; detection of motion artifacts and contact loss; and separation of cervical EMG from EEG using spectral and learned methods suited to few-channel recordings. Outputs will include the comparison metrics in Table 1, computed reproducibly for every layout and material tested. The pipeline will be released as open-source code.

Aim 3b: Estimating sleep information from patch-compatible signals

Public PSG datasets (Kemp et al., 2000; Goldberger et al., 2000; Zhang et al., 2018) will be used to train models restricted to signals that approximate what the patch could record: a single EEG derivation near the mastoid, ECG, and an EMG-derived muscle-tone signal. Features will include EEG band power and spindle and slow-wave measures; heart rate, RMSSD, and HF and LF power in windows centered on each 30-s epoch; and EMG power as an index of muscle tone. Two model classes will be compared: a feature-based gradient-boosted classifier and a deep sequence model (Supratak et al., 2017; Perslev et al., 2021). Ablation analyses will compare EEG-only, cardiac-only, and combined models, with particular attention to REM sleep.

Aim 3c: Transfer to measured data

Models from Aim 3b will be applied to the nap recordings from Aims 1b and 2c, with and without fine-tuning, to estimate how well they transfer to signals recorded at the patch site. Because afternoon naps contain little REM sleep, REM-specific evaluation in this phase relies on the public datasets, and transfer analyses focus on wake, N1, and N2.

Validation and outcomes

All evaluation will use subject-wise cross-validation so that no participant contributes data to both training and testing. Outcomes will include epoch-by-epoch Cohen’s kappa relative to expert scoring, interpreted using conventional benchmarks (Landis and Koch, 1977); per-stage sensitivity, specificity, and F1 score, with emphasis on REM sleep; and the change in performance from adding cardiac and muscle-tone features to EEG.

Potential problems and alternative approaches

Public PSG signals only approximate patch signals; this will be addressed by adding realistic noise and artifact profiles measured in Aims 1 and 2 to the public data. Class imbalance, particularly for N1, will be addressed with class-weighted loss functions and per-stage reporting. If five-stage performance is limited, a four-class scheme (wake, light NREM, deep NREM, REM) commonly used for wearable devices will also be reported.

Rigor and Reproducibility

Quality criteria, primary outcomes, and analysis plans will be preregistered before data collection. Expert sleep scoring will be blinded to patch-site data. Sex will be considered as a biological variable in all analyses. De-identified data and analysis code will be shared publicly where participant consent permits.

Human Subjects Protection and Ethical Considerations

All procedures will be reviewed and approved by an appropriate Institutional Review Board before any participant is enrolled, and written informed consent will be obtained from all participants. Anticipated risks are minimal and include skin irritation from electrodes, materials, or adhesives; transient discomfort; and mild sleep disturbance during nap sessions. These risks will be mitigated by advancing only materials with established skin-contact biocompatibility data, by inspecting the skin before and after each session, and by allowing participants to remove any material or withdraw at any time without penalty. Data will be de-identified at collection and stored on encrypted, access-controlled systems. All test articles are investigational research materials and are not intended for diagnosis. If a recording suggests a possible clinically relevant abnormality, such as an irregular heart rhythm, the participant will be informed and advised to consult a physician.

Timeline

The R&D phase is planned over 18 months, with the three aims running in parallel (Table 2).

Table 2. Timeline and deliverables by aim.
ActivityMonthsDeliverables
Ethics approval and preregistration1–3Approved protocol; preregistered criteria
Aim 1a: Anatomical review and modeling1–5Ranked layout shortlist; first manuscript draft
Aim 1b: Measurement study5–11Selected layout; signal-quality report
Aim 2a: Bench screening of materials2–9Material shortlist; biocompatibility summary
Aim 2b: Skin-wear testing8–12Skin tolerance and adhesion results
Aim 2c: Small-format test patches12–16Test-patch recordings; comparison with conventional electrodes
Aim 3a: Signal-quality and artifact pipeline1–8Open-source analysis pipeline
Aim 3b–c: Sleep information estimates and transfer4–18Modality ablation results; transfer analysis
Integration16–18Prototype design specification; R&D manuscript

Expected Outcomes and Deliverables

  • Anatomical and signal evidence. Quantitative evidence on where and how EEG, ECG, and EDA can be recorded at the retroauricular–cervical site, including the influence of muscle, hair, skin properties, and pressure.

  • Materials shortlist. A set of breathable, skin-safe substrates, electrodes, and adhesives with measured electrical stability and skin tolerance.

  • Analysis pipeline. Open-source tools for signal-quality scoring, artifact separation, and sleep feature extraction.

  • Sleep information estimates. Estimates of how well sleep stages, including REM sleep, can be identified from patch-compatible signals, and of the contribution of each signal.

  • Prototype design specification. An integrated specification of layout, materials, and analysis approach for the first full prototype, with its known limitations documented.

Potential Applications

  • REM sleep tracking. A patch that records brain activity, heart rhythm, and muscle tone at one site could track REM sleep across many nights in participants’ own homes, supporting research on REM sleep, dreaming, and memory that is currently limited by the burden of PSG.

  • More comfortable clinical sleep studies. Clinical sleep studies require many wired sensors on the scalp, face, and body, which many patients find uncomfortable and which can disrupt the sleep being measured. A validated patch could reduce the number of sensors needed for some clinical questions, enable more comfortable multinight or home-based monitoring, and be especially useful for patients who tolerate full PSG poorly. Any clinical use would require separate validation and regulatory review.

  • Funding pathway. Because this project has a potential product outcome, demonstrated feasibility would create a clearer route to industry sponsorship or venture investment than purely theoretical research, which could reduce dependence on grant funding as development progresses. Any such development would be managed under institutional conflict-of-interest policies.

Future Directions

Following this R&D phase, the design specification will be used to build an integrated wireless prototype. That prototype will then be validated overnight against simultaneous PSG in a larger sample using subject-independent models, with full-night REM sleep detection as a primary outcome, and evaluated during multinight home use. Later studies may extend testing to older adults and to clinical populations in which comfortable sleep monitoring would be most valuable.

Limitations

Several limitations are anticipated. First, this phase studies healthy adults aged 18–45 years, so results may not generalize to older adults or clinical populations. Second, measurements are made during daytime sessions and naps rather than full nights, which limits direct assessment of N3 and REM sleep. Third, estimates from public PSG datasets rely on signals that only approximate those recorded at the patch site. Fourth, a retroauricular–cervical site provides limited coverage of frontal and occipital cortex and lacks dedicated EOG, and EDA at a nonpalmar site may have reduced sensitivity. These limitations will be reported explicitly and will guide the design of the subsequent validation phase.

Conclusion

This proposal outlines the initial research and development needed to design a comfortable, breathable skin patch for multimodal sleep monitoring. By studying the anatomy of the recording site, screening skin-safe materials, and building quantitative tools before committing to a full device, this phase will determine what a single patch can realistically measure and how it should be built. The resulting design specification will support subsequent development for REM sleep tracking and for more comfortable clinical sleep studies.

Team

Research team

  • Cristobal Sanchez — Lead Theorist & Project Director
  • Pranathi Raghavendra — Head of Anatomy & Physiology
  • Jin-Rou (Zoey) Chen — Biochemical Interactions Lead
  • Millan Hamilton — Computational Analysis Lead

Faculty AdvisorAnonymous, Affiliation withheld

References

References

  1. [1]Benedek M, Kaernbach C (2010) A continuous measure of phasic electrodermal activity. J Neurosci Methods 190:80–91.
  2. [2]Berry RB, Quan SF, Abreu AR, et al. (2020) The AASM manual for the scoring of sleep and associated events: rules, terminology and technical specifications, version 2.6. Darien, IL: American Academy of Sleep Medicine.
  3. [3]Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310.
  4. [4]Boucsein W (2012) Electrodermal activity, 2nd ed. New York: Springer.
  5. [5]Buysse DJ, Reynolds CF 3rd, Monk TH, Berman SR, Kupfer DJ (1989) The Pittsburgh Sleep Quality Index: a new instrument for psychiatric practice and research. Psychiatry Res 28:193–213.
  6. [6]Chinoy ED, Cuellar JA, Huwa KE, et al. (2021) Performance of seven consumer sleep-tracking devices compared with polysomnography. Sleep 44:zsaa291.
  7. [7]de Zambotti M, Cellini N, Goldstone A, Colrain IM, Baker FC (2019) Wearable sleep technology in clinical and research settings. Med Sci Sports Exerc 51:1538–1557.
  8. [8]Debener S, Emkes R, De Vos M, Bleichner M (2015) Unobtrusive ambulatory EEG using a smartphone and flexible printed electrodes around the ear. Sci Rep 5:16743.
  9. [9]Dong L, Zhang X, Li M, et al. (2026) AI-enabled wearable ultrasound for noninvasive central venous pressure monitoring. Cyborg Bionic Syst 7:0653.
  10. [10]Goldberger AL, Amaral LAN, Glass L, et al. (2000) PhysioBank, PhysioToolkit, and PhysioNet: components of a new research resource for complex physiologic signals. Circulation 101:e215–e220.
  11. [11]Goncharova II, McFarland DJ, Vaughan TM, Wolpaw JR (2003) EMG contamination of EEG: spectral and topographical characteristics. Clin Neurophysiol 114:1580–1593.
  12. [12]Hobson JA, Pace-Schott EF, Stickgold R (2000) Dreaming and the brain: toward a cognitive neuroscience of conscious states. Behav Brain Sci 23:793–842.
  13. [13]Huang Y, Parra LC, Haufe S (2016) The New York Head—a precise standardized volume conductor model for EEG source localization and tES targeting. Neuroimage 140:150–162.
  14. [14]Kemp B, Zwinderman AH, Tuk B, Kamphuisen HAC, Oberyé JJL (2000) Analysis of a sleep-dependent neuronal feedback loop: the slow-wave microcontinuity of the EEG. IEEE Trans Biomed Eng 47:1185–1194.
  15. [15]Kim D-H, Lu N, Ma R, et al. (2011) Epidermal electronics. Science 333:838–843.
  16. [16]Klem GH, Lüders HO, Jasper HH, Elger C (1999) The ten-twenty electrode system of the International Federation. Electroencephalogr Clin Neurophysiol Suppl 52:3–6.
  17. [17]Landis JR, Koch GG (1977) The measurement of observer agreement for categorical data. Biometrics 33:159–174.
  18. [18]Looney D, Kidmose P, Park C, et al. (2012) The in-the-ear recording concept: user-centered and wearable brain monitoring. IEEE Pulse 3:32–42.
  19. [19]Mikkelsen KB, Villadsen DB, Otto M, Kidmose P (2017) Automatic sleep staging using ear-EEG. Biomed Eng Online 16:111.
  20. [20]Miyamoto A, Lee S, Cooray NF, et al. (2017) Inflammation-free, gas-permeable, lightweight, stretchable on-skin electronics with nanomeshes. Nat Nanotechnol 12:907–913.
  21. [21]Pan J, Tompkins WJ (1985) A real-time QRS detection algorithm. IEEE Trans Biomed Eng 32:230–236.
  22. [22]Perslev M, Darkner S, Kempfner L, et al. (2021) U-Sleep: resilient high-frequency sleep staging. NPJ Digit Med 4:72.
  23. [23]Sano A, Picard RW, Stickgold R (2014) Quantitative analysis of wrist electrodermal activity during sleep. Int J Psychophysiol 94:382–389.
  24. [24]Shaffer F, Ginsberg JP (2017) An overview of heart rate variability metrics and norms. Front Public Health 5:258.
  25. [25]Somers VK, Dyken ME, Mark AL, Abboud FM (1993) Sympathetic-nerve activity during sleep in normal subjects. N Engl J Med 328:303–307.
  26. [26]Sterr A, Ebajemito JK, Mikkelsen KB, et al. (2018) Sleep EEG derived from behind-the-ear electrodes (cEEGrid) compared to standard polysomnography: a proof of concept study. Front Hum Neurosci 12:452.
  27. [27]Supratak A, Dong H, Wu C, Guo Y (2017) DeepSleepNet: a model for automatic sleep stage scoring based on raw single-channel EEG. IEEE Trans Neural Syst Rehabil Eng 25:1998–2008.